180
_____________________________________________________________________________________________
10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:WITHstand:CLOWer <data>
<data>
ON/OFF
(Character data)
Function
Enables and disables the lower-limit value for withstand-voltage tests, in the READY state.
Example
Enables the lower-limit value for a withstand-voltage test.
Transmission :WITHstand:CLOWer ON
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
Syntax
:WITHstand:CLOWer?
Response
syntax
<data>
Function
Queries the lower-limit value enablement for withstand-voltage tests.
Note
ON/OFF
(Character data)
Example
Queries the lower-limit value enablement for withstand-voltage tests.
Transmission :WITHstand:CLOWer?
Response ON
Syntax
:WITHstand:TIMer <data>
<data>
ON/OFF
(Character data)
Function
Enables and disables the test time for withstand-voltage tests, in the READY state.
Example
Enables the test time for a withstand-voltage test.
Transmission :WITHstand:TIMer ON
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
:WITHstand:CLOWer
Enables and disables the lower-limit value for withstand-voltage tests.
:WITHstand:CLOWer?
Queries the lower-limit value enablement for withstand-voltage tests.
:WITHstand:TIMer
Enables and disables the test time for withstand-voltage tests.

Table of Contents