10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:WITHstand:TIMer?
Response
syntax
<data>
<data>
ON/OFF
(Character data)
Function
Queries the test time enablement for withstand-voltage tests.
Example
Queries the test time enablement for withstand-voltage tests.
Transmission :WITHstand:TIMer?
Response ON
Syntax
:WITHstand:UTIMer <data>
<data>
ON/OFF
(Character data)
Function
Sets the ramp-up time ON/OFF for withstand-voltage tests in the READY state.
Example
Sets the ramp-up time for the withstand-voltage tests to ON.
Transmission :WITHstand:UTIMer ON
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
Syntax
:WITHstand:UTIMer?
Response
syntax
<data>
<data>
ON/OFF
(Character data)
Function
Queries the ramp-up time ON/OFF for withstand-voltage tests.
Example
Queries the ramp-up time ON/OFF for withstand-voltage tests.
Transmission :WITHstand:UTIMer?
Response ON
:WITHstand:TIMer?
Queries the test time enablement for withstand-voltage tests.
:WITHstand:UTIMer
Sets the ramp-up time ON/OFF for withstand-voltage tests.
:WITHstand:UTIMer?
Queries the ramp-up time ON/OFF for withstand-voltage tests.