EasyManua.ls Logo

Hioki 3153 - Page 208

Hioki 3153
255 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
195
_____________________________________________________________________________________________
10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:MEMory:INSulation:SAVE <data>
<data>
1to10
(NR1)
Function
Saves the current insulation-resistance test settings in the memory file specified in <data>.
Example
Saves insulation-resistance test data to memory file 2.
Transmission :MEMory:INSulation:SAVE 2
Error
The execution of this command in a state other than the insulation-resistance mode READY state
causes an execution error.
The execution of this command in the withstand-voltage mode and program mode also causes an
execution error.
An execution error occurs when file numbers other than 1 to 10 are specified.
Syntax
:MEMory:INSulation:CLEar <data>
<data>
1to10
(NR1)
Function
Clears and resets the data saved for the insulation-resistance test specified in <data>.
Example
Clears the insulation-resistance test data saved to memory file 3.
Transmission :MEMory:INSulation:CLEar 3
Error
The execution of this command in a state other than the READY state causes an execution error.
An execution error occurs when file numbers other than 1 to 10 are specified.
:MEMor
y
:INSulation:SAVE
Saves data for insulation-resistance tests to memory files.
:MEMor
y
:INSulation:CLEar
Clears the insulation-resistance test data saved to memory files.

Table of Contents

Related product manuals