10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:PROGram:EDIT:STEP <data1>,<data2>,<data3>,<data4>,<data5>,<data6>,<data7>,
<data8>,<data9>,<data10>,<data11>,<data12>,<data13>,<data14>
<data1>
1to32
:Filenumber
<data2>
1to50
: Step number
<data3>
0
(OFF),
1to4
: Scanner box number (high voltage side)
<data4>
1to4
(Multiple mode),
1to8
(Single mode): Scanner channel number (high voltage side)
<data5>
0
(OFF),
1to4
: Scanner box number (low voltage side)
<data6>
5to8
(Multiple mode),
0
(Single mode): Scanner channel number (low voltage side)
<data7>
0
: Stop upon completion of this step.,
1
: Proceed to the next step.
<data8>
Withstand-voltage test
Test type:
0
(AC50 Hz-withstand-voltage test),
1
(AC60 Hz-withstand-voltage test),
2
(DC-withstand-voltage test)
<data9>
Test-voltage value:
0.20 to 5.00
<data10>
Upper-limit value:
0.1 to 100
(AC-withstand-voltage test),
0.1 to 10
(DC-withstand-voltage test)
<data11>
Lower-limit value:
0
(OFF),
0.1 to 99
(AC-withstand-voltage test),
0.1 to 9.9
(DC-withstand-voltage test)
<data12>
Test time:
0.3 to 999
<data13>
Ramp-up timer:
0
(OFF),
0.1 to 99.9
<data14>
Ramp-down timer:
0
(OFF),
0.1 to 99.9
<data8>
Insulation-resistance test
Test type:
3
(insulation-resistance test)
<data9>
Test-voltage value:
50 to 1200
<data10>
Upper-limit value:
0
(OFF),
0.10 to 9999
<data11>
Lower-limit value:
0.10 to 9999
<data12>
Test time:
0.3 to 999
<data13>
Delay time:
0
(OFF),
0.1 to 99.9
0
(OFF)
<data14>
Input "0" at all times (<data 1> is always in NRf format).
Function
Sets the scanner, test mode, and test conditions for the desired file and step for the program tests.
Example
In step 2 of program test file No. 1, a withstand-voltage test is conducted with CH 1 of scanner
box 1 on the high-pressure side and CH 7 of scanner box 2 on the low-pressure side, before
proceeding to the next step. (The scanner is set in multi-mode.)
Makes settings as follows: File No. 1, step No. 2, CH 1 of scanner box 1 (high pressure), CH 7
of scanner box 2 (low pressure), proceeds to the next step, AC 50 Hz withstand-voltage test, test
voltage 1.50 kV, upper-limit value 10 mA, lower -limit value OFF, test time 5.0 s, ramp-up time
2.0 s, ramp-down time OFF
Transmission :PROGram:EDIT:STEP 1,2,1,1,2,7,1,0,1.50,10,0,5.0,2.0,0
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
The transmission of the following settings causes an execution error:
Insulation-resistance mode
(1) Delay time
≧
test time
(2) Due to the relationship between the test voltage and the lower-limit value, the transmission of
a test voltage or lower-limit value outside the range causes an execution error.
:PROGram:EDIT:STEP
Makes settings for the scanner, test mode, and test conditions for program test step.