10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:PROGram:EDIT:STEP? <data1>,<data2>
<data1>
1to3
(NR1): File number
<data2>
1to50
(NR1): Step number
Response
syntax
<data1>,<data2>,<data3>,<data4>,<data5>,<data6>,<data7>,<data8>,<data9>,<data10>,
<data11>,<data12>,<data13>,<data14>
<data1>
1to32
(NR1): File number
<data2>
1to50
(NR1): Step number
<data3>
0
(OFF),
1to4
: Scanner box number (high voltage side) (NR1)
<data4>
1to4
(Multiple mode)
1to8
(Single mode): Scanner channel number (high voltage side)
(NR1)
<data5>
0
(OFF),
1to4
: Scanner box number
<data6>
5to8
(Multiple mode),
0
(Single mode): Scanner channel number (low voltage side) (NR1)
<data7>
0
: Stop upon completion of this step.,
1
: Proceed to the next step. (NR1)
<data8>
Withstand-voltage test
Test type:
0
(AC50 Hz-withstand-voltage test),
1
(AC60 Hz-withstand-voltage test),
2
(DC-withstand-voltage test)
<data9>
Test-voltage value:
0.20 to 5.00
(NR2)
<data10>
Upper-limit value:
0.1 to 100
(AC-withstand-voltage test),
0.1 to 10
(DC-withstand-voltage test)
(NR1 or NR2)
<data11>
Lower-limit value:
0
(OFF),
0.1 to 99
(AC-withstand-voltage test),
0.1 to 9.9
(DC-withstand-
voltage test) (NR1 or NR2)
<data12>
Test time:
0.3 to 999
(NR1 or NR2)
<data13>
Ramp-up time:
0
(OFF),
0.1 to 99.9
(NR1 or NR2)
<data14>
Ramp-down time:
0
(OFF),
0.1 to 99.9
(NR1 or NR2)
<data8>
Insulation-resistance test
Test type:
3
(insulation-resistance test) (NR1)
<data9>
Test-voltage value:
50 to 1200
(NR1)
<data10>
Upper-limit value:
0
(OFF),
0.10 to 9999
(NR1 or NR2)
<data11>
Lower-limit value:
0.10 to 9999
(NR1 or NR2)
<data12>
Test time:
0.3 to 999
(NR1 or NR2)
<data13>
Delay time:
0
(OFF),
0.1 to 99.9
(NR1 or NR2)
<data14>
0
(OFF) (NR1)
Function
Queries the scanner, test mode, and test conditions of the file and step to be confirmed for
program tests.
Example
Queries the scanner mode of program test file No. 1.
Transmission :PROGram:EDIT:STEP? 1,2
Response 1,2,1,1,2,7,1,0,1.50,10,0,5.0,2.0,0
File No. 1, step No. 2, CH 1 of scanner box 1 (high pressure), CH 7 of
scanner box 2 (low pressure), proceeds to the next step, AC 50 Hz withstand-
voltage test, test voltage 1.50 kV, upper-limit value 10 mA, lower-limit value
OFF, test time 5.0 s, ramp-up time 2.0 s, ramp-down time OFF
:PROGram:EDIT:STEP?
Queries settings for the scanner, test mode, and test conditions for program test step.