199
_____________________________________________________________________________________________
10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:PROGram:LOAD:FILE <data>
<data>
1to32
(NR1)
Function
Loads the program test file.
Example
Loads program test file No. 5.
Transmission :PROGram:LOAD:FILE 5
Error
The execution of this command in a state other than the READY state or "test setting state"
causes an execution error.
An execution error occurs when file numbers other than 1 to 32 are specified.
Syntax
:MEASure:RESult:FILE?
Response
syntax
<data>
<data>
0
(PASS) or
1to50
(Step No. s that resulted in a FAIL) (NR1)
Function
Queries the test result of the program test file that executed the test.
Example
Queries the test result of the program test file that executed the test.
Transmission :MEASure:RESult:FILE?
Response 5
The result was FAIL in step 5.
Note
An execution error will occur if testing is terminated before all steps that have been programmed
have ended.
:PROGram:LOAD:FILE
Loads the program test file.
:MEASure:RESult:FILE?
Queries the test result of the program test file.