10.3 RS-232C Command Reference
______________________________________________________________________________________________
Syntax
:MEASure:RESult:STEP? <data>
<data>
1to50
(NR1)
Response
syntax
<data1>,<data2>,<data3>,<data4>,<data5>,<data6>
(
<data6>
: withstand-voltage test only)
When the designated step is a withstand-voltage test
<data1>
Test type:
0
(AC50 Hz-withstand-voltage test),
1
(AC60 Hz-withstand-voltage test),
2
(DC-
withstand-voltage test) (NR1)
<data2>
Measured voltage value (unit: kV)
<data3>
Measured current value (unit: mA)
<data4>
Test time elapsed (unit: s)
<data5>
Determination:
PASS
UFAIL
(UPPER FAIL)
LFAIL
(LOWER FAIL)
ULFAIL
(UPPER LOWER FAIL)
OFF
(Others) (Character data)
<data6>
Timer type:
0
(test time),
1
(ramp-up time),
2
(ramp-down time) (NR1)
Test type, determination, Other than the type of timer: NR1 or NR2
When the designated step is an insulation-resistance test
<data1>
Test type:
3
(insulation-resistance test) (NR1)
<data2>
Measured voltage value (unit: V)
<data3>
Measured resistance value (unit: MΩ)
O.F.: 10000
U.F.: 0
<data4>
Test time elapsed (unit: s)
<data5>
Determination:
PASS
UFAIL
(UPPER FAIL)
LFAIL
(LOWER FAIL)
ULFAIL
(UPPER LOWER FAIL)
OFF
(Others) (Character data)
Test type, excluding determination: (NR1 or NR2)
Function
Queries the measured result in the step.
Example
Queries the measured result in step No. 3.
Transmission :MEASure:RESult:STEP? 3
Response 3,500,120,10.0,PASS
Output voltage 500 V, measured resistance value 120 M
Ω
, test time 10.0 s,
test result PASS.
:MEASure:RESult:STEP?
Queries the measured result in the program test step.