6 Calibration Verifications
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 106
4. If needed, access the test strip provider. Remove and discard all
chemistry test strips present inside the chamber; see Accessing the
Strip Provider Module.
5. Pull the Strip Loader out of the
system. Load the CalChek strips
inside the Strip Loader, and then
push the Strip Loader back inside
the system. Rotate the Strip
Loader 180 degrees to transfer
the strips to the Strip Provider
Module.
6. Verify that the Lot ID number and the expiration date from the
CalChek strip container match the data from the screen. Modify if
necessary.