8 Maintenance & Service
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 144
Monthly Maintenance
WARNING: Wear protective gloves to prevent exposure to pathogens.
Discard contaminated materials according to applicable regulations.
Cleaning the Wash Station Bath
Items required: Deionized water, cotton swabs and protective gloves
1. Make sure the System is in Standby
mode, as indicated on the top left of the
instrument screen.
2. Click on the Instrument button on the top right side of the main screen.
3. Click on the Go off line button. The system status will change to Off
Line.
4. Turn the power off by pressing
the green button located on the
left side of the chemistry
system.
5. Open the front door to access
the wash station bath.
6. Using a cotton swab moistened
with deionized water, remove
salt deposits present on the
wash bath.
CAUTION: Do not insert the cotton swab inside the wash tube; cotton
particles may clog the tubing connectors.
7. Close the front door.
8. Turn the power on by pressing the green button located on the left
side of the chemistry system.
9. Click on the Go On line button. The system status will change to On
Line.