10 Consumables Traceability
301-7146 English Rev B 11/02/2011 Operators Manual NA 204
Alarm 52
Causes Remedies
This alarm was triggered because
the IRISpec CC™ control material
is expired.
This condition should be remedied
by running a control rack with a
new IRISpec CC™ control material.
Alarm 54
Causes
Remedies
This alarm was triggered because
chemistry strip is expired.
This condition should be remedied
by running a control or patient rack
with new chemistry strips.
Alarm 59
Causes
Remedies
This alarm was triggered because
the IRISpec CA™ control material
is expired.
This condition must be remedied
for the instrument to run more
specimens.
Run a control rack with a new
IRISpec CA™ control material
or
Check the Expired Consumable
Lockout Override checkbox in
the QC setting window. See
Expired Consumable Lockout
Override.