7 Results
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 127
2. Refill the sample tube(s) with at least 2.0 mL of specimen.
3. Re-run the specimen(s).
Processing not complete
The system has detected an internal irrecoverable error.
The instrument stops processing new specimens and raises an audible
and/or visual alarm to signal a problem, and displays a message
indicating that an unrecoverable error has occurred.
The system flags the remaining specimens in process with “Processing
not complete” flags and for each one indicates the specimen ID, rack ID,
tube position, Operator, and Flag on the Work list.
Remedies
1. Reject the flag to remove the results from the Work List.
2. Refill the sample tube(s) with at least 2.0 mL of specimen.
3. Re-run the specimen(s).
S1
The instrument cannot self-calibrate with the iChem Wash Solution.
Call your technical support representative for further assistance.
S2
The system has detected an internal irrecoverable error pertaining to the
Color Clarity Specific Gravity module.
Call your technical support representative for further assistance.
S3
The Specific Gravity result is >1.060. This is a normal condition.
1.000
The Specific Gravity result is <1.000. This is a normal condition.