1 Introduction
301-7146 English Rev B 11/02/2011 Operators Manual NA 21
Requirements
The system has few special environmental requirements. It uses
alternating current at 100V to 240V and 50 Hz to 60 Hz. (Input voltage
and frequency selection does not require customer intervention.)
Uninterruptible power supplies are recommended for the iChem
VELOCITY to maintain system operation during short power outages and
brownouts. This allows for an orderly shutdown of instruments without the
loss of data.
Select a room to set up the unit where the temperature can be controlled
between 68
o
F (18
o
C) and 82
o
F (28
o
C), and relative humidity non-
condensing in a range of 20% to 80%.
CAUTION: The equipment inside fixed covered areas of the instrument is
not operator serviceable and may present electrical shock hazards.
WARNING: For continued protection against risk of fire and hazard,
replace only with the same type and rating fuse.
WARNING: Equipment requires connection to protective earth for safety
reasons.
WARNING: The instrument’s main supply inlet is being used as the
mains disconnect device.
7BInstallation
The iChem VELOCITY will be installed by a factory-trained representative
from Iris Diagnostics or an authorized Distributor.