9 iWARE
Expert System
301-7146 English Rev B 11/02/2011 Operators Manual NA 177
Chapter 9 - iWARE
Expert System
Data Management
The Data Management function uses specific filtering rules set up by the
laboratory. When microscopy and/or chemistry results become available,
the filtering rules are applied. Any result matching a rule can be used to
signal for review by the operator.
The Data Management applies the rules on each sample at the end of
each run and takes appropriate action:
Set a Validation flag and prevent auto-release
Output additional data to LIS
Signal the operator in the Information pane during sample review
to review the sample
Utilize reporting tags for screen and printer based reports
This function allows the user:
To create, edit, and save logic statements for filtering results
based upon multiple Urine Chemistry parameters, multiple
Microscopy parameters, and patient demographics (gender, age,
location).
To apply to each parameter, a user-defined value and a logic sign.
To recreate pre-populated logic statements.
To display results on-screen and generate printable work lists
based on logic statements.
To easily activate / disable every saved logic statement without
deleting.
The Data Management function has three components:
1. Rule Generator
2. Results/Rules Tagging
3. Report Generator