7 Results
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 132
5. Select the new result from the drop down list. The new value is
highlighted in yellow, indicating that a change was made. If
necessary, check the new value for abnormal result.
6. When all results have been entered, enter a comment in the
Comment field according to the Laboratory Protocol.
7. Enter the analyzer identification from which the chemistry results were
obtained.
8. Click the OK button to validate the changes and update the results
report.
Entering Manually a Microscopy Result Value
This option can be used to enter manually microscopy results to be
associated to specific chemistry results,
1. From the Worklist screen, double click the specimen row or select the
row and click the Specimens button. The Specimen screen is displayed.
2. From the Specimen screen, click the Edit button. The Edit Which
Results window is displayed.