8 Maintenance & Service
301-7146 English Rev B 11/02/2011 Operators Manual NA 160
Alarm Messages
Alarm 17
Causes
Remedies
The instrument is getting close to
having too many unreleased results
in memory.
Handle specimens on the Work
List, which will free some memory.
Alarm 18
Causes
Remedies
The instrument has too many
unreleased results in memory.
This condition must be remedied
for the instrument to run more
specimens.
Handle specimens on the Work
List, which will free some memory.
Alarm 19
Causes
Remedies
The system could not communicate
with the LIS. These are the
possible causes:
The LIS cable has been
disconnected.
The LIS is down.
The LIS is too busy to respond
(overloaded).
Reconnect the LIS cable.
Check with your LIS manager
to see when the LIS will be
back up.
Request increased LIS capacity
from your LIS manager.