8 Maintenance & Service
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 137
Daily Maintenance
WARNING: Wear protective gloves to prevent exposure to pathogens.
Discard contaminated materials according to applicable regulations.
Empty and Clean the Waste Container
Items required: Iris System Cleanser diluted 1:10 (REF 475-0003),
paper towels, and protective gloves.
Empty the waste container after every 300 strips
1. Open the waste container door.
2. Remove the waste container and
discard used test strips according to
local regulations.
3. Clean the waste container with Iris
System Cleanser diluted 1:10 (REF
475-0003), and then rinse with tap
water.
4. Thoroughly dry the waste container.
5. Reinstall the waste container.
6. Close the waste container door.
Discarding Liquid Waste
WARNING: Dispose of waste product, unused product and contaminated
packaging in compliance with applicable legal regulations. If unsure of
the applicable legal requirements, contact the local authorities for
information.
This procedure is not required if the instrument is connected to a
drain.
Items required: Mild detergent, protective gloves.
1. Remove the drain bottle cap and tubing from the drain bottle.
2. Discard the liquid waste.