10 Consumables Traceability
301-7146 English Rev B 11/02/2011 Operators Manual NA 205
Alarm 60
Causes Remedies
This alarm was triggered because
the IRISpec CB™ control material
is expired.
This condition must be remedied
for the instrument to run more
specimens.
Run a control rack with a new
IRISpec CB™ control material
or
Check the Expired Consumable
Lockout Override checkbox in
the QC setting window. See
Expired Consumable Lockout
Override
.
Alarm 61
Causes Remedies
This alarm was triggered because
the IRISpec CC™ control material
is expired.
This condition must be remedied
for the instrument to run more
specimens.
Run a control rack with a new
IRISpec CC™ control material
or
Check the Expired Consumable
Lockout Override checkbox in
the QC setting window. See
Expired Consumable Lockout
Override
.