4 Sample Processing
301-7146 English Rev B 11/02/2011 Operators Manual NA 81
Entering Manual Orders
1. Select the first rack number to be used.
2. Enter the specimen information:
a. Specimen Identifier
Patient ID can be entered in this field. If a barcode label is
present on the tube, the results will be displayed with the
information from the barcode. The Work Order option will
change to "Run".
b. Fluid Type - select URN.
c. Dilution - not available for the iChemVELOCITY.
d. Work Order - default No Order
No Order
The sample will be processed according to the user
defined criteria.
3. When the specimen information is entered, place the sample tube into
the corresponding position of the selected sample rack.
4. Repeat for each specimen to be run for the selected sample rack.
5. If more than one rack is to be run, select a new rack number and then
enter the specimen information.
6. Click OK to save the manual entries only or click OK & Print to save
the entries and print a report.