EasyManua.ls Logo

IRIS iChemVelocity

IRIS iChemVelocity
238 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8 Maintenance & Service
301-7146 English Rev B 11/02/2011 Operators Manual NA 169
Switching from Backup to Primary Chemistry System
To return the primary system back online, perform the following steps to
switch to the primary chemistry system.
NOTE: Only a Manager can modify the settings. Any user can view the
settings.
1. On the iChem Boost box, switch the knob located on the front to the A
position.
2. From the Instrument screen, click on the Go off line button. Click Yes
3. Click on the Settings button located at the bottom of the Instrument
screen. The Settings screen is displayed.
4. From the Setting screen, click on the System Configuration button.
The System Configuration screen is displayed.

Table of Contents