8 Maintenance & Service
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 163
Alarm 40
The Reflectance Calibration Check
failed.
Repeat the Reflectance Calibration
Check. If it continues to fail, please
call your technical support
representative for further
assistance.
Alarm 41
The SG/Color/Clarity Calibration
Check failed.
Repeat the SG/Color/Clarity
Calibration Check. If it continues to
fail, please call your technical
support representative for further
assistance.
Alarm 42
The chemistry system’s wash
solution level is low. It is depleted
in the normal process of running
the system.
Replace the wash solution.
Alarm 43
The chemistry system is out of test
strips. The test strips are depleted
in the normal course of running the
system. Alternatively, a strip may
be jammed in the strip provider
module.
If the strip feeder is empty add
fresh strips. Otherwise check the
strip provider module for jams, see
Strip Provider Module Jam.
Empty the chemistry strip waste
container.