8 Maintenance & Service
301-7146 English Rev B 11/02/2011 Operators Manual NA 167
Switching from Primary to Backup Chemistry System
If the primary system is not working,
perform the following steps to switch to a
backup chemistry system.
An iChem BOOST box allows the user to
enable the backup chemistry system.
NOTE: Only a Manager can modify the settings. Any user can view the
settings.
1. On the iChem Boost box, switch the knob located on the front to the B
position.
2. From the Instrument screen, click on the Go off line button. Click Yes
3. Click on the Settings button located at the bottom of the Instrument
screen. The Settings screen is displayed.
4. From the Setting screen, click on the System Configuration button.
The System Configuration screen is displayed.