5 Quality Control
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 92
Reviewing Quality Controls Results
1. Click Instrument.
2. Click Quality Review. The Quality Review screen is displayed.
3. Data can be sorted by clicking on a column header. A repeat click will
toggle between ascending (∆) and descending (∇) order. The default
is Date/Time – ascending order.
4. The Quality Control report can be reviewed by highlighting the desired
ID and by clicking on Re-report. A screen is displayed where the QC
report destination can be selected (i.e. printer, LIS, or both). The
report will consist of Control name, system identifier, analysis time
stamp, report time stamp, rack position, lot ID, Control expiration date,
test strip lot ID, test strip expiration date and status (pass/fail). The
report will also indicate the results for all the analytes, lower and
upper limit of acceptance and status (pass/fail). The overall status will
be indicated as fail if any of the analytes has failed.