10 Consumables Traceability
301-7146 English Rev B 11/02/2011 Operators Manual NA 207
Alarm 68
Causes Remedies
This alarm was triggered because
Reflectance Check control material
is expired.
This condition should be remedied
by running a control rack with a
new Reflectance Check control
material.
Alarm 69
Causes Remedies
This alarm was triggered because
Reflectance Check control material
is expired.
This condition must be remedied
for the instrument to run more
specimens.
Run a control rack with a new
Reflectance Check control
material or
Check the Expired Consumable
Lockout Override checkbox in
the QC setting window. See
Expired Consumable Lockout
Override
.
Expired Consumable Lockout Override
Only a manager can enable the ‘Expired Consumable Lockout Override’
function for a sample to be run on the instrument. This checkbox is
enabled only if an expired consumable material is used.
When a manager overrides the lockout, the Expired Consumable red
alarm becomes a yellow alarm and the ‘Expired Consumable Lockout
Override’ checkbox is disabled again until an expired consumable
material is used.
Accessing the Expired Consumable Lockout Override
Option
1. Click on the Instrument button located on the top right side of the
main screen.