11 Appendix
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 218
Instrument to instrument precision
Three levels of target analyte concentrations for each assay (negative,
first positive level and highest positive level) were tested 60 times on
each of three iChemVELOCITY systems. The exact % agreement with
expected results is shown.
Bilirubin
Urobilinogen
Ketones
Ascorbic acid
Glucose
Protein
Blood
pH
Nitrites
Leukocytes
*Subtilisin is a surrogate for leukocyte esterase. 3.41 uL/mL subtilisin is
equivalent to 25 WBCs/mL and 50 μL/mL is equivalent to 500 WBCs/μL.