8 Maintenance & Service
301-7146 English Rev B 11/02/2011 Operators Manual NA 138
3. Clean the drain bottle with a mild detergent and water, and then rinse
with tap water.
4. Insert the drain tube into the drain bottle and tighten the cap.
Cleaning the Sample Transport Module
Items required: Iris System Cleanser diluted 1:10 (REF 475-0003), lint-
free tissue and protective gloves.
1. Moisten a tissue with Iris System Cleanser diluted 1:10 (REF 475-
0003) and wipe the Sample Transport Module to remove any
deposits. Check under the belts and the pulleys.
2. Wipe again using distilled water.
3. Wipe dry.
Cleaning the Optional Load/Unload Trays
Items required: Iris System Cleanser diluted 1:10 (REF 475-0003), lint-
free tissue and protective gloves.
1. Moisten a tissue with Iris System Cleanser diluted 1:10 (REF 475-
0003) and wipe the Load/Unload Stations to remove any deposits.
Check under the belts and the pulleys.
2. Wipe again using distilled water.
3. Wipe dry.
Cleaning the Instrument Surfaces
Items required: Iris System Cleanser diluted 1:10 (REF 475-0003), lint-
free tissue and protective gloves.
1. Moisten a tissue with Iris System Cleanser diluted 1:10 (REF 475-
0003) and wipe the instrument to remove any deposits.
2. Wipe again using distilled water.
3. Wipe dry.