7 Results
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 126
3. The confirmation results may be added in the Edit Comment box, if
desired.
ID_Error
The barcode reader was not able to read the barcode label.
Remedies
1. From the Work List screen, double-click on the specimen to open the
Specimen Review screen.
2. Click on the Review Flag Specimen button, the Edit Specimen
Identifier screen is displayed.
3. Enter the specimen ID in the Specimen identifier edit box.
4. Click OK. The screen closes and the Specimen screen displays the
new specimen ID.
5. Click Accept to clear the ID flag and refresh the screen.
Processing Failure
The system has detected an internal error. If the error is recoverable, the
instrument continues processing strips that are already dosed.
Remedies
1. Reject the flag to remove the results from the Work List.