11 Appendix
301-7146 English Rev B 11/02/2011 Operators Manual NA 215
iChemVELOCITY Urinalysis System
Method Comparison
The performance of the iChemVELOCITY was compared with
commercially available urinalysis systems. Urine samples were from a
hospital or a commercial laboratory. Samples with concentrations
covering the entire assay reporting range were tested. The urinalysis
systems used provided semi-quantitative and/or qualitative results with
the exception of SG, where quantitative results were provided. The exact
% agreement and % agreement within +/- one report level is shown for all
tests. SG comparison was done by linear regression (n= 248, R2 =
0.999, slope = 1.013, intercept = -0.014).
Test
Number
of
Samples
% Exact
Agreement
% +/- One Report
Agreement
Bilirubin
416
99.5
100
Urobilinogen
453
98.9
100
Ketones
480
98.9
100
Ascorbic acid
416
99.3
100
Glucose
446
97.5
100
Protein
490
98.6
100
Blood
446
97.5
100
pH
96
93.8
100
Nitrites
374
99.5
100
Leukocytes
494
98.5
100
Expected Values and Ranges
See package insert for more information.