10 Consumables Traceability
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 206
Alarm 63
Causes Remedies
This alarm was triggered because
chemistry strip is expired.
This condition must be remedied
for the instrument to run more
specimens.
• Run a control or patient rack
with new chemistry strips or
• Check the Expired Consumable
Lockout Override checkbox in
the QC setting window. See
Expired Consumable Lockout
Alarm 66
This alarm was triggered because
SG/Color Clarity control material is
expired.
This condition should be remedied
by running a control rack with a
new SG/Color Clarity control
material.
Alarm 67
This alarm was triggered because
SG/Color Clarity control material is
expired.
This condition must be remedied
for the instrument to run more
specimens.
• Run a control rack with a new
SG/Color Clarity control
material or
• Check the Expired Consumable
Lockout Override checkbox in
the QC setting window. See
Expired Consumable Lockout