10 Consumables Traceability
301-7146 English Rev B 11/02/2011 Operators Manual NA 206
Alarm 63
Causes Remedies
This alarm was triggered because
chemistry strip is expired.
This condition must be remedied
for the instrument to run more
specimens.
Run a control or patient rack
with new chemistry strips or
Check the Expired Consumable
Lockout Override checkbox in
the QC setting window. See
Expired Consumable Lockout
Override.
Alarm 66
Causes
Remedies
This alarm was triggered because
SG/Color Clarity control material is
expired.
This condition should be remedied
by running a control rack with a
new SG/Color Clarity control
material.
Alarm 67
Causes
Remedies
This alarm was triggered because
SG/Color Clarity control material is
expired.
This condition must be remedied
for the instrument to run more
specimens.
Run a control rack with a new
SG/Color Clarity control
material or
Check the Expired Consumable
Lockout Override checkbox in
the QC setting window. See
Expired Consumable Lockout
Override.