9 iWARE
Expert System
301-7146 English Rev B 11/02/2011 Operators Manual NA 185
a. Use the drop-down button to select the first criteria for the rule.
Formed particles for microscopy,
Analytes for chemistry,
Last name, first name, age, gender, or location for
demographics.
b. Use the drop-down button to select the logic sign. The following
are available:
= (Equal to)
< (Less than)
> (Greater than)
<= (Less than
or equal to)
>= (Greater than
or equal to)
!= (Not equal to)
c. Use the drop-down button to select the specific value. Values are
pre-populated from the user-defined settings. See Chapter 3
Setup. If the drop-down button is not available, use the keyboard
to enter a specific value.
Adding More Conditions to a Rule
1. To enter more conditions, click the Add Condition button. The Add
Condition screen is displayed.
2. Select the condition number before which a new condition will be
inserted. Only one condition can be selected. Click the OK button.
3. Using the drop-down button, select the first criteria for the new
condition: AND/OR; and then repeat step 5 from Adding one
Condition. Below is an example of a Rule Condition.
Deleting a Condition from a Rule
1. To delete a condition, click the Delete Conditions button. The Delete
Conditions screen is displayed.
2. Select the condition(s) to suppress, and then click the OK button.

Table of Contents