10 Consumables Traceability
301-7146 English Rev B 11/02/2011 Operator’s Manual NA 196
3. To access the first chemistry control level, click on Next.
4. Enter the name of the control material, lot ID, and expiration date.
Select the lower and upper limits for each analyte listed.
NOTE: The lower and upper limits can be selected by using the drop-
down buttons only.
5. Color, Clarity, and Ascorbic Acid are not applicable for CA and CB.
Ascorbic Acid is applicable for CC.
6. To access the next control level, click Next. (delete)
7. Repeat the process for the second control, using the next button.
8. Enter the acceptable range for each analyte for that control.