8.3 Acoustic Stando󰀨
268
8
Appropriately t Acoustic Stando󰀨 so that
it is positioned as shown in the figures
below.
9
Put an appropriate amount of ultrasound
gel on the Acoustic Standoff area on the
lens surface of the transducer, and start
an examination.
z
After use
1
Detach Acoustic Stando󰀨 from the trans-
ducer.
Slowly detach it in reverse order of when at-
tached. The ultrasound gel may be spattered.
2
Wipe the ultrasound gel o󰀨 the transducer
and Acoustic Standoff, and clean / disin-
fect each of them in turn according to the
instructions provided in "8.3.3 Cleaning
and Disinfection of Acoustic Stando󰀨".
3
Wipe sterile water off Acoustic Standoff
with a clean cloth, and let it dry.
4
Cover Acoustic Standoff with a pair of
protective films, and put / store it in the
storage bag.
Do not leave it on paper, plastic, etc.
Avoid storing Acoustic Stando󰀨 in a place where
it can be exposed to direct sunlight, radiation
such as X-rays, and high temperature / humidity,
and store it in a dark place.
z
Disposal
Please follow the rules and regulations of your relevant
authorities in the disposal of an Acoustic Stando󰀨 and
its packing materials.
(The materials used for the Acoustic Stando󰀨 are am-
mable and do not leave any hazardous residue to the
environment after being burned.)
The Acoustic Standoff is categorized as industrial
waste. If you discard without being sterilized after use,
it could be categorized as specially controlled industrial
waste. Be sure to follow the rules and regulations of
the local governing body, and request an approved in-
dustrial waste disposal company to discard the Acous-
tic Stando󰀨.