27
DANGER
To prevent the risk of injury or explosion, do not
operate this system in a ammable or oxygen-en-
riched atmosphere. Be sure to strictly observe the
following precautions. Otherwise the battery may
leak, overheat, or rupture.
Do not expose the battery to re or heat it.
Do not allow the terminals to be connected by
metal or conducting material.
The battery must be charged in the system.
Do not dismantle or modify the battery.
Do not apply solder directly to the battery.
The installation orientation of the battery is speci-
ed. When connecting the battery to the system,
be sure not to connect it with the terminals re-
versed.
Do not connect the battery to a wall mounted AC
socket outlet or a car cigarette lighter socket.
Do not use the battery with systems other than
this system.
If liquid leaking from the battery comes into con-
tact with your eyes, ush them immediately with
clean water and consult a physician. Do not rub
your eyes. Failure to follow these measures may
result in blindness.
Do not use this system for an eyeball examination
or any other examination that may require the pas-
sage of an ultrasound beam through an eyeball.
This system is not designed to be applied to eye-
balls.
Do not use this system for intracardiac use or direct
cardiac contact. It may injure patients.
WARNING
Do not disassemble, repair, or modify the main unit.
Doing so may cause an electric shock.
To perform repair, special techniques are re-
quired. Contact your service representative.
Do not insert or remove the mains plug with wet
hands. It can cause an electric shock.
Do not connect the power supply cord of this
system to a wall mounted AC socket outlet which
is connected to other important medical devices
through the same circuit breakers and fuses. In the
event of this system malfunctioning, a current surge
may occur resulting in other medical devices mal-
functioning, which may injure patients.
Keep the power supply cord and mains plug safe
from being damaged. For example, do not scratch,
alter, place near a heater, forcibly bend, twist, pull,
place a heavy object on or bundle the power sup-
ply cord and mains plug. Use of a damaged power
supply cord or mains plug can result in an electric
shock, short circuit or re.
Contact your service representative for repair of
the power supply cord and mains plug.
Do not place devices around the mains plug. It may
pose a risk to the patient as the power is not turned
OFF when needed.
Do not remove the covers or panels of the system,
as electric shock may result.
When the system is powered ON, the operator
must not touch any exposed connectors. Doing so
may cause an electric shock.
Keep liquids away from the system, options or pe-
ripheral devices. Otherwise there is a risk of electric
shock.
When using other equipment such as an electric
scalpel, high-frequency therapy equipment or elec-
tric stimulator, make sure to keep patients away
from the ultrasound transducer. The patient may
get burned or receive an electric shock.