WARNING
• Be sure to use the power supply cord that is in-
cluded with the system, and connect the mains
plug (Hospital Grade) to a wall mounted AC socket
outlet. Do not connect the mains plug to a two-pin
socket outlet using conversion adapters, etc. Doing
so may cause an electric shock.
• Where the integrity of the external protective con-
ductor of the facility or its arrangement is doubt-
ful, the system shall be operated with the battery-
powered operation.
• Do not connect a multiple socket-outlet or an exten-
sion cord between the mains plug (with three-core
ground terminal) and a hospital grade mains socket
outlet (three-pin). Doing so may cause an electric
shock.
• Securely insert the mains plug completely in. In-
complete insertion may cause electric shock or re
through overheat.
– Do not use damaged plugs or loose socket out-
lets.
– When the mains plug is not in use, pull it out from
the wall mounted AC socket outlet.
• Routinely remove dust or dirt settling on the mains
plug. Dust or dirt accumulation on the mains plug
may result in an insulation failure, causing a re.
– Disconnect the mains plug and wipe it clean with
a dry cloth.
• To prevent electric shock, do not connect any pe-
ripheral units other than external monitor to an ex-
ternal outlet. Peripheral units should be connected
to the service outlet of the Power Extension Unit2.
For the connection procedures, contact your ser-
vice representative.
• Do not connect peripheral units other than AC
adapter or printer authorized by Konica Minolta to
the Power Extension Unit2. Doing so may reduce
the safety of the system.
• When the potential equalization terminal on this
system is connected to the potential equalization
busbar of the facility via a potential equalization
conductor, the potential of connected equipment or
of a connected system is equalized.
WARNING
Reference Signal Unit
• To prevent the electric shock, be sure to conrm the
following cautions before use.
– There should be no abnormality on the ECG cable.
– ECG Unit should be connected to the specified
ECG cable for the Reference Signal Unit.
• Connect the Reference Signal Unit to the system
prior to attach the ECG electrodes to a patient.
Also, avoid bringing the patient with the ECG elec-
trodes attached into contact with other conductive
parts including grounding. Doing so may cause an
electric shock to the patient when the connector of
the ECG cable touches the live part.
• Do not connect the Reference Signal Unit to any
connector other than USB connector. Doing so may
cause a breakdown.
• Do not let ECG electrodes to touch patient’s heart
directly. Doing so may cause a cardiac arrest.
– ECG is Type BF applied part, and is not intended
for direct cardiac application.
• Be sure to use the ECG cable that is included with
the Reference Signal Unit.
• The selection and use of ECG electrodes are the
responsibility of a user.
• Remove any ECG electrodes from your patient be-
fore using equipment such as an electric scalpel,
high-frequency therapy equipment, electric stimulator
or electric
debrillator. Also, make sure to keep any
ultrasound transducer away from the patient. The pa
-
tient may get burned or receive an electric shock
.
Battery
• Do not immerse the battery in water or seawater,
and do not allow the terminal sections to become
wet. Otherwise, the battery may overheat or the ter-
minal sections may rust.
• Do not remove the external cover from the battery
or damage the cover. Doing so may cause the bat-
tery to leak or overheat.
• If leakage, discoloration, deformation, or other ab-
normalities are found in the battery, stop using the
battery.
• If charging fails to complete after the specied re-
charging time has elapsed, immediately stop further
recharging. Otherwise, the battery may leak, over-
heat, or rupture.
• If liquid leaking from the battery comes into contact
with skin or clothing, ush the aected area imme-
diately with clean water. Otherwise, skin problems
may occur.
• The battery has a limited lifetime. If the operating
time is shorter even when the battery has been fully
charged under the specied conditions, replace the
battery with a new one.