9/13/00 Printing a Report 8-9
Summary Report Template (SSA instrument)
Report Settings
Lower Level
Report Settings Description
Unformated [No] A dump of all
Read
,
Query
and
Interva
l remote commands.
Multiple values are comma delimited.
Data Report [Yes] This shows general information such as Leq, SEL, Dose, Projected Dose,
Lmin, Lmax, Peak, Ln’s, exceedance, and calibration logs.
Setup Report [No] Print the settings and their values for the SSA instrument.
Intv Report [No] Leq, SEL, Lmin, Lmax, Peak, Uwpk, exceedances, time, date, and dura-
tion in the short report. The Long reports adds the Ln values and Spectral
LNs for the Interval.
Hist Report [No] Leq for each time history period.
Print Range Intv Report [No] You can customize the interval report by choosing the long or short
report.
Lowest Record (00000) Selects the lowest interval record to be printed.
Highest Record (65535) Selects the highest interval record to be printed.
Hist Report [No] You can customize the time history report by choosing the long or short
report.
Lowest Record (00000) Selects the lowest time history record to be printed.
Highest Record (>99999) Selects the highest time history record to be printed.
Histograms SPL Histogram [No] Prints SPL histogram and overall Spectral LN report.
Lowest Level (045) Selects the lowest SPL histogram level to be printed.
Highest Level (095) Selects the highest SPL histogram level to be printed.
Resolution (5.0) Selects the SPL histogram resolution
Pk-1 Histogram [No] You can customize the Peak1 Histogram report by choosing the long or
short report.
Lowest Level (110) Selects the lowest Peak1 histogram level to be printed.
Highest Level (160) Selects the highest Peak1 histogram level to be printed.
Resolution (5.0) Selects the Peak1 histogram resolution
Pk-2 Histogram [No] You can customize the Peak2 Histogram report by choosing the long or
short report.
Lowest Level (070) Selects the lowest Peak2 histogram level to be printed.