B-16 824 Reference Manual 9/12/00
Sound Exposure (E) is calculated and displayed in pascal
squared hours as needed in some of the European Countries.
This value is read with the R162 command and is available
in the top line of the DOSE-a display. The formula used is:
E = T • (10^(Leq/10)) • 20µPa
2
,
where E is the Exposure, T is the elapsed time in hours, and
20µPa
2
is the reference sound pressure 20 micro pascal
squared.
R154 ALL Internal UwPk detector level nnn.nn dB
R155 ALL Number of Bytes in Keyboard Stack
This is used with keyboard simulation
to prevent the loss of keys.
nnn
R156 LOG Background Leq Time hhhhh:mm:ss.ss
R157 ALL Calibration Status Done, Wait, Waiting, Stable,
Unstable, Too High, Too
Low.
R158 ALL Total Memory Available nnnnnn
R159 ALL Memory Available in percent nnn.nn
R160 ALL OVERALL OVERLOAD FLAG c
R161 RTA Number of RTA Records nnnnn
R162 LOG Overall Sound Exposure nnnnn.n P
2
H
R163,x SSA, ISM,
LOG
TAKT 3 nnn.n dB
R164,x SSA, ISM,
LOG
TAKT 5 nnn.n dB
R165 LOG PassBy Time of Occurrence ddmmmyy hh:mm:ss
R166 LOG PassBy Event Duration hh:mm:ss / mm:ss.ss

Table of Contents

Other manuals for Larson Davis System 824

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Larson Davis System 824 and is the answer not in the manual?

Larson Davis System 824 Specifications

General IconGeneral
BrandLarson Davis
ModelSystem 824
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals