B-50 824 Reference Manual 9/12/00
Brackets in the syntax indicate optional characters and oper-
ands.
The time syncing feature can be used to synchronize the
interval history records to the instruments real-time clock. If
the Interval Period is set to the time shown in the first col-
umn below, the first interval of each new measurement will
end at the real time indicated in the second column (the hh
means any hour and the hh:m means any hour and tens of
minutes). All subsequent intervals will now be synchro-
nized to the real-time clock:
Syntax Response
I[nterval]var_no Intv_var
Examples:
I 1
I 1,-5
I1,5
01Jan1997 01:25:00
01Jan1997 01:20:00 (from 5 previous)
01Jan1997 01:30:00 (from 5 after)
Intv
Period
Sync On Sync to nearest
01:00 hh:00 hour
00:30 hh:00, hh:30 half hour (30 minutes)
00:20 hh:00, hh:20, hh:40 one-third hour (20 minutes)
00:15 hh:00, hh:15, hh:30, hh:45 one-fourth hour (15 minutes)
00:10 hh:m0 one-sixth hour (10 minutes)
00:05 hh:m0, hh:m5 one-twelfth hour (5 minutes)
00:0x hh:mm minute

Table of Contents

Other manuals for Larson Davis System 824

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Larson Davis System 824 and is the answer not in the manual?

Larson Davis System 824 Specifications

General IconGeneral
BrandLarson Davis
ModelSystem 824
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals