9/12/00 Overview 2-15
Step 3
To setup the calibration process, press the
r
key
to move into the Calibration Menu.
Step 4
To set the calibration level, press the
d
and
u
keys to highlight “Cal Level”,
Step 5
To set the value of the sound level provided by the
calibrator, press the
r
key. This displays the
Cal
Level
window.
Step 6
Then use the
l
and
r
keys (to move the cursor
position) and the
u
and
d
keys (to change dig-
its) to enter the proper level.
Step 7
To select the new calibration level press the
c
Check key.