2-16 824 Reference Manual 9/12/00
Step 8
To verify that the instrument is properly cali-
brated, use the
u
key to highlight “Check”.
Step 9
Place the calibrator over the microphone, switch
the calibrator ON, and press
c
to activate the cal-
ibration
Check
process.
The display will show the measured level and the difference
from the calibrator level entered and the measured level as
shown.
Step 10
If the calibration is not reasonably close to the
level of the calibrator, the calibration can be
changed by choosing
Change
. Press
c
, use
the
d
key to highlight “Change”.
Step 11
Switch the calibrator ON, and press
c
.