V Series Operating Instructions 23 - 7
Alarm Watch Alarm Watch Notification
23.3.1 Acknowledging Watched Alarms
To acknowledge a single alarm (where the Show All button is not present):
• Select the A-W tile or the Alarm Watch Notification Strip (as shown in FIGURE 23-5).
This opens the Remote View dialog and silences all the active alarms for the bed.
NOTE: Refer to the Remote View chapter for additional information.
To acknowledge an alarm where the Show All button is present:
1. Select the Show All button.
The Alarm Watch Overview dialog (shown in FIGURE 23-6) displays.
2. Select the Acknowledge All button to acknowledge all the alarms at the Host monitor for the
monitored beds or wait for the dialog to time-out.
FIGURE 23-6 Example All Current Alarms Dialog
NOTE: The Alarm Watch Overview dialog displays the highest priority most
recent alarms for the currently monitored beds.