EasyManua.ls Logo

Minebea Intec PR5900 - Page 109

Default Icon
382 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WP-A
Max 3000g d= 0.1g
00
+
g
Calibration @admin
1
2
3
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040800
1.177298
0.987577
0.470919
30000 d
1 d
103.966 g
3000.000 g
2516.5 g
98.11 cnt/d
Save
For new calibrations, the parameters [Scale interval], [Dead load at] and [Max at] are
inactive.
Note:
If a linearization was carried out (see Chapter 4.7.1.2.1.4.4), the following message
appears once the parameter has been selected: "Cannot be changed while
linearization is active."
Only deleting of the linearization points deactivates the linearization mode!
[Save] is deactivated for new calibrations. The next menu item [Scale interval] is
activated.
Possible error messages, see Chapter 9.7.1.
WP-A
Max 3000g d= 0.1g
00
+
g
Calibration @admin
1
2
3
Max
Scale interval
Dead load at
Max at
calibrated at
Sensitivity
3000.0 g
0.1 g
mV/V
mV/V
mV/V
µV/d
0.040368
1.177688
0.987884
0.471075
30000 d
1 d
102.832 g
3000.000 g
2516.5 g
98.14 cnt/d
Save
STOP
Set Max failed.
> phys. Max
Ok
If [OK] is pressed, the input value for the maximum capacity is canceled.
1. Select [Max] and activate the input eld.
2. Enter the value with decimal places (in this example: 3000.0).
3. Press the key to select the weight unit.
4. Conrm the entries.
The conrmation is displayed with "setting Max…".
5. Press the [Save] soft key to save the calibration.
The reference value is calculated for the electronics test.
4 System setup Process Controller Maxxis 5 PR 5900
EN-107 Minebea Intec

Table of Contents

Related product manuals