- Serial number of xBPI scale or weighing module, if used in legal metrology
Accessible via MENU - [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [xBPI scale] - [Parameters] .
WP-C
Max 29.9kg d= 0.0005kg
00000
kg
Weighing points @admin
Weighing point A
Weighing point B
Weighing point C
Weighing point D
Internal weighing point
Internal weighing point
xBPI scale
Liquid counter
Parameters Calib Units Setup
WP-C
Max 29.9kg d= 0.0005kg
00000
kg
Weighing points @admin
Weighing point C
Settings locked
W&M
Tare timeout
WP serial number
SBN address
xBPI scale
none
2.0 s
0
0
Default Save
[Settings locked]
The weighing point settings and assignment can only be viewed as if the
corresponding CAL switch were closed; see Chapter 2.7.1.
[W&M]
Setting for operation in legal metrology.
Selection: [None], OIML, NTEP (for USA) or NSC (for Australia), see Chapter 4.7.1.1.1.
[Tare timeout]
Timeout for a zeroset or tare command to be executed. If the xBPI scale has not
executed the command in the specied time, the action will be aborted.
Setting range: 0…9.9 s
[WP serial number]
Serial number of the connected xBPI scale/weighing module. The number is required
for checking when used in legal metrology. With serial number 0, checking is omitted.
1. In this case: Select [Weighing point C] with [xBPI scale] and press the [Parameters]
soft key.
A selection window opens.
2. Select and conrm the individual parameters.
4 System setup Process Controller Maxxis 5 PR 5900
EN-135 Minebea Intec