- [Parameters], see Chapter 4.7.8.2.
- [Calib], see Chapter 4.7.8.3.
- [Units] (display accuracy), see Chapter 4.7.8.4.
The digital load cells have been calibrated at the factory based on the acceleration of
gravity at Hamburg (9.81379 m/s
2
). The calibration data in the load cells are invariable.
The calibration data for the gravity acceleration at the place of installation can be adapted
only in the device and protected against overwriting (see Chapter 2.7). With applications
for use in legal metrology, the legal requirements and the conditions given on the test/
approval certicate must be taken into account when selecting the settings.
Accessible via MENU - [Operating] - [System setup] - [Weighing points] - [Weighing
point x] - [Weighing point] - [Pendeo Process] .
Weighing point D @admin
Weighing point not assigned
Save
Weighing point D @admin
Weighing point not assigned
Save
Not assigned
Internal weighing point
Liquid counter
User scale
xBPI scale
SBI scale
Pendeo Truck
Pendeo Process
PR-Net weighing point
Mettler-Scale
SMA scale
1. Conrm [Weighing point].
A selection window opens.
2. Select and conrm [Pendeo Process].
3. Press the [Save] soft key to save the settings.
If a scale has not yet been connected, the message "scale not ready" appears.
The weighing electronics are now assigned to the weighing point.
Process Controller Maxxis 5 PR 5900 4 System setup