WP-C
Max -------?? d= -------??
no values from scale
Weighing points @admin
Weighing point A
Weighing point B
Weighing point C
Weighing point D
Internal weighing point
Internal weighing point
SMA scale
Pendeo Truck
Parameters
WP-C
Max -------?? d= -------??
no values from scale
Weighing points @admin
Weighing point C
Settings locked
W&M
Max
Scale interval
Minimum weight
SMA scale
none
1000 kg
1
0 d
Default Save
[Settings locked]
The weighing point settings and assignment can only be viewed as if the
corresponding CAL switch were closed; see Chapter 2.7.1.
[W&M]
Setting for operation in legal metrology.
Selection: [None], OIML, NTEP (for USA) or NSC (for Australia), see Chapter 4.7.1.1.1.
[Max] (maximum capacity)
See Chapter 4.7.1.2.1.1.
[Scale interval]
See Chapter 4.7.1.2.1.2.
[Min. weight]
Minimum weight at which a print command can be triggered.
Setting range: 0…9999999 d.
For use in legal metrology a value of at least 20 d must be set.
1. In this case: Select [Weighing point D] with [SMA scale] and press the [Parameters]
soft key.
A selection window opens.
2. Select and conrm the individual parameters.
Process Controller Maxxis 5 PR 5900 4 System setup
Minebea Intec EN-224