WP-A
Max 3000g d= 0.1g
00
–
g
Weighing points @admin
Settings locked
W&M
Measurement time
Digital filter
Cutoff frequency
External load cell supply
none
160 ms
aperiodic
2.00 Hz
below or equal 8 V
Default Save
[Settings locked]
The weighing point settings and assignment can only be viewed as if the
corresponding CAL switch were closed; see Chapter 2.7.1.
[W&M]
Setting for operation in legal metrology.
Selection: [None], OIML, NTEP (for USA) or NSC (for Australia)
For description and settings, refer to Chapter 4.7.1.1.1
[Measurement time]
The duration of a measurement can be selected.
Selection: 5 ms, 10 ms, 20 ms, 40 ms, 80 ms, 160 ms, 320 ms, 640 ms, 960 ms,
1280 ms, 1600 ms, [≤1 s]
A measurement time of ≤1 s must be selected for use in legal metrology.
[Digital lter]
The following includes examples of interference signals for the dierent lter types:
- Bessel lter
2. Select and conrm the individual parameters.
Process Controller Maxxis 5 PR 5900 4 System setup
Minebea Intec EN-94