Technical data
Description Data
Connection Terminal, 7-pin
Number of outputs 3 (CH1, CH2, CH3)
Max. switching voltage 31 V DC
Protection against incorrect polarity.
Max. switching current 25 mA
Voltage drop @ 25 mA: 3 V
Passive, external power supply required.
Potential isolation Via optocoupler.
Cables Screened
Connect cable screen (wire gauge max. 1.5 mm
2
) to the termi-
nal (terminal contact 7; see example).
Cable length Max. 50 m
X3 Process Indicator PR 5410 4 Device installation
Minebea Intec EN-42