EasyManua.ls Logo

Olympus FLUOVIEW FV300 - Page 135

Olympus FLUOVIEW FV300
564 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPLIED OPERATIONS/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2-21
Page
10 Setting a Lower Scan Sp eed
1. The scan speed can be decreased using the scale in the [Scan Speed] group box
on the center of the [Acquire] panel.
In general, setting a lower scan speed allows the acquired image quality to
be improved.
However, a low scan speed also lengthens the time required for image
acquisition.
When the scan speed is decreased during fluorescence observation,
the saturation of fluorescence may darken the image of certain types
of specimens. In this case, increase the scan speed and increase the
PMT Voltage or use accumulation in scanning.
TIP
NOTE
[Scan Speed] group box
Set the scan speed by clicking a point on the
scale line.

Table of Contents

Other manuals for Olympus FLUOVIEW FV300

Related product manuals