EasyManua.ls Logo

Olympus FLUOVIEW FV300 - Editing the Protocol

Olympus FLUOVIEW FV300
564 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPLIED OPERATIONS/Protocol processor
IV.
OPERATION INSTRUCTIONS
IV .
2-151
Page
: <Get FV Status> button: Acquires each setting value of FLUOVIEW
software parameters and registers to PAPP window. The track which the
parameters to be registered has to be highlighted.
: <Set to FV> button: Sets PAPP parameters to FLUOVIEW software
parameters.. The track which the parameters to be registered has to be
highlighted.
<Get FV Status> and <Set to FV> button can not be pressed during
PAPP protocol is working.
2-4-2 Editing the Protocol
The protocol of the Protocol Processor consists of several Tracks, and Track is a set of
image acquisition condition and parameter settings.
The [PAPP] window is used to edit the protocol.
The Track needs to be edited as a set of acquisition condition, then deferent Track has
to be created in case of the condition becomes different as shown below;
Track 1
Interval:
30sec(3)
A
cquisition every 3sec(1):
4 frames(2)
Track 2
Frame Frame Frame Frame Frame Frame
3sec. 100sec.
A
cquisition every 100sec(4):
2 frames(5)
For details, see section 2-4-8,” Example of assembling a protocol”.
(1)
(2)
(4)
(5)
(3)

Table of Contents