Getting Started FLUOVIEW/Online Help
IV.
OPERATION INSTRUCTIONS
IV
. 1-56
Page
One Point!
Select the <Contents> button to the initial display. Select the <Back> button to return to
the previous information page.
1-3-2 Setup of Microscope and Scan Unit
The microscope and scan unit can be set up from the FLUOVIEW software, by selecting
the observation method and following the displayed guidance information.
x Selecting the Dyeing Method
1. From the page tabs on the bottom right of the [Acquire] panel, select the [Dyes]
sub-panel.
Fig. 1-8 [Dyes] Sub-panel
2. Select the specimen dyeing method by dragging desired dye names in the
[Available Dyes] list box in the [Selected Dyes] group box to the field immediately
above the list box.
3. Click the <Apply> button to apply the selected dyeing method to the [Channel
1]/[Channel 2] group box on the upper part of the [Acquire] panel.
[Available Dyes] list box
Lists the available dyes. Select the
desired items from this list and drag
them to the field above it to select the
d
y
ein
g
method.
<Clear> button
Clear the set dyeing method.
<Prev.> button
Sets the dyeing method which
was set last time by clicking the
<Apply> button.
<Apply> button
A
pplies the dyeing method
dragged in the [Selected
Dyes] group box to the
[Channel 1 ]/ [Channel2]
group in the [Acquire] panel.
Place the pointer on the icon
displayed in the [Selected
Dyes], and the dyeing method
is shown in the pop-up
display.
[Assign dyes manually] check box
Checking this enables the manual
setting. Dragging the dyeing method
in the list directly to the [Ch] group box
assigns the dye to the desired
channel.