Getting Started FLUOVIEW/Outline of LSM Observation Procedures
IV.
OPERATION INSTRUCTIONS
IV.
1-29
Page
If you want to use a differential interference unit in transmitted light
observation, refer to the instruction manual of your microscope.
With transmitted light differential interference
observation using an immersion objective, set the
microscope’s field diaphragm so that it circumscribes
the field of view. Otherwise the contrast may degrade.
1-2-3-2 Combination with IX
1. Turn the light path selector (1) on the right side of the microscope to
(when using the Manual microscope)
From the page tabs on the bottom right of the [Acquire] panel, select the
[Settings] sub-panel(when using the Motorized microscope).
Fig. 1-4 [Settings] Sub-panel
2. Select the <BI> button in the [Light Path] group box (when using the
Motorized microscope).
The <BI> button looks pushed in to indicate that it is selected.
3. Engage the optimum cube for specimen dye by operating the cube turret
(2) (when using the Manual microscope)
Engage the optimum cube for specimen dye by pressing the cube button
(3) on the hand switch (4).
NOTE
(3)
(4)
(1)