Getting Started FLUOVIEW/Outline of LSM Observation Procedures
IV.
OPERATION INSTRUCTIONS
IV
. 1-48
Page
1-2-12-7 Setting the Multiple sections to be Observed
While acquiring image, move the Z stage to select the multiple sections to be observed.
From the panel page tabs shown on the bottom right of the [Acquire] panel, select the [Z
Stage] sub-panel.
1. Check the [Locked] check box in the [Z Stage] sub-panel.
Do not turn the fine focus adjustment knob while the [Locked] check
box is checked, for this may damage the Z motor.
2. While observing the image in the [Live] panel, locate the plane to be observed by
displacing the stage using the <Z stage coarse adjustment> and <Z stage fine
adjustment> buttons in the [Z Stage] sub-panel.
[Current Pos] text box
Shows the current position of the stage.
The value can also be entered directly
from the keyboard.
[Locked] check box
Z-motor is engaged by checking this
box.
<Z stage fine adjustment> buttons
Displace the Z stage by 0.1 Pm per
step.
<Z stage coarse adjustment>
buttons
Displace the Z stage by 1.0 Pm
per step.